Planning a Concentration Series

How to space concentrations for an experiment (linear, 2-fold, half-log or 10-fold), when to dilute directly or serially, how to keep pipetting volumes accurate, and why the vehicle should stay constant.

Reviewed 20 Oct 2026 2 references Companion tool: T-12 Concentration worksheet

Many experiments test a compound at several concentrations. A well-planned series covers the right range with sensible spacing, keeps every pipetting step accurate, and keeps everything except the compound constant.

Choosing the spacing

Spacing Factor between points Points per 10-fold Good for
2-fold 2 ≈ 3.3 Detailed curves over a narrow range
Half-log √10 ≈ 3.16 2 Dose–response curves over several decades
10-fold 10 1 Range finding, first experiments
Linear Equal steps — Standard curves within a linear range

Dose–response behaviour is usually logarithmic, so equal ratios (2-fold, half-log) spread points evenly across a sigmoid curve. Linear spacing suits calibration curves where the response is linear in concentration.

Direct or serial?

  • Direct dilution makes each tube from the same stock. Errors do not accumulate, but the lowest concentrations need very small stock volumes.
  • Serial dilution makes each tube from the previous one. Volumes stay comfortable, but pipetting errors compound down the series.
  • Intermediate stocks combine both: dilute the stock once or twice to intermediates, then make each working solution directly from the nearest intermediate.

Keep pipetting volumes in range

For each tube, the stock volume is:

Vstock = Ctarget × Vfinal / Cstock

The worksheet flags any transfer below 2 µL, because volumes that small are hard to deliver accurately with standard pipettes. Fix flagged rows by making an intermediate stock (for example 10-fold) and recalculating those tubes from it, or by increasing the final volume.

Worked example

Stock 10 mM, final volume 1 mL, targets 100, 50, 25, 10, 5 and 1 µM.

Stock volumes: 10, 5, 2.5, 1, 0.5 and 0.1 µL. The last three are below 2 µL. Make a 100 µM intermediate (10 µL of stock + 990 µL of diluent), then prepare 10, 5 and 1 µM from it with 100, 50 and 10 µL in 1 mL.

Keep the vehicle constant

If the stock is in DMSO or another co-solvent, diluting it directly gives every tube a different solvent concentration. Solvents can affect cells and assays on their own, so a concentration effect becomes a solvent effect. Two common fixes:

  • Make the dilution series in the co-solvent first, then add the same small volume of each to the assay buffer, so every tube gets the same solvent fraction.
  • Add solvent to the lower-concentration tubes so they match the highest one.

Include a vehicle-only control at the same solvent concentration.

Before you start

  • Check that the highest target is below the stock concentration and within solubility in the final buffer.
  • Prepare 10–20% more volume than you need, to allow for pipetting losses.
  • Print the worksheet, tick off each tube as you go, and keep it with your notebook.